| Company | Products | Lab | Technical Library | Training | What's New | Contact Us |

 

Infrared Calibration
Blackbody Sources Hyperion R Model 982

Hyperion R

Low Temperature Blackbody Source, -10°C to 80°C E >0.995

  DB4From Databook 4
  • Low Temperature Radiation Pyrometer Primary Source
  • 50mm Cavity
  • 0.995 Emissivity
Data Sheet



The Hyperion R Calibration Bath comprises a blackbody cavity for calibration of infrared thermometers. An important factor is that its operation range includes ambient temperature, which is a difficult radiation point to achieve.

The temperature of the furnace is set on a controller, while an independent indicator, whose sensor is built into the cavity, indicates the actual radiance temperature. The temperature of the cavity can be set to within 0.01°C at any point from -10°C to +80°C.

This elegant, modern-styled range has been designed to work from a series of Peltier modules, which can be used either to heat or to cool a metal block. In conjunction with a specially programmed microprocessor-based controller it offers a unique concept in calibration sources.

With this bath, a temperature above, below or at ambient temperature can be set and will be maintained whether the source needs to be cooled or heated. Traceability may be established with a UKAS certificate for the in-built indicator and supplied probe (935-14-13).

© 2008 Isothermal Technology Ltd. All rights reserved.

Home Page Primary Lab Secondary Lab Industrial Calibration InfraRed Calibration