The Medusa R comprises a blackbody with which radiation pyrometers can be calibrated.
The temperature of the cavity can be set to ±0.1ºC over the temperature range 30ºC to 550ºC.
The controller reaches 550ºC approximately 45 minutes from switch on. Also, the cavity temperature, as
measured by a radiation thermometer took approximately 45 minutes to reach stability from switch on.
The
variation seen on the controller’s temperature indication over a 5 minute period was ±0.2ºC. Similar
variations were detected by a radiation thermometer looking into the cavity.
Using a portable radiation thermometer having a target diameter of 13mm, the 45mm target was surveyed.
Maximum temperature differences of ±1ºC were found for set temperatures in the range 100ºC to 500ºC.
Temperatures along the inner 200mm of the 285 mm long cavity were measured at 400ºC and 500ºC, using
a hand held fibre-optic radiation probe. Maximum temperature differences of ±4ºC were found.
The temperature, as shown on the controller, agrees with the cavity temperature as measured by a
radiation thermometer, where calibration is traceable to National Standards, to within ±2ºC.
Slim Fixed
Point Cells can be fitted to the Medusa R giving small uncertainties suitable for National
Standards.
Traceability may be established with a UKAS certificate for the in-built indicator and supplied probe (935-14-95). |