Cal Notepad

Log and Display Isotech Indicators and Baths

Included with Isotech Equipment

 

 


  • Easy to use Win98 - Win XP
  • Interface to Isotech Block Baths & Temperature Indicators
  • Log Chart and Export data - Control Calibration Bath - Read Standard
Data Sheet

Purpose

The purpose of Cal NotePad (CNP) is to automatically log and display the temperature of an Isotech calibration bath together with the unit under test. Cal NotePad is included with Isotech Equipment.

Cal Notepad can be used with baths (or indicators) without PC interfaces by the user typing in values from the keyboard. The Cal Notepad can be used to identify the operator and the unit under test. With the click of a button data is logged with time information, it is also possible to log continually.

The calibration bath temperature can be changed from the PC or from the calibration baths controller - Cal Notepad will display the temperature changes as they occur on the re-scaleable chart display.

Traceable Calibration

For traceable calibration the unit under test should be compared to a calibrated standard thermometer. Cal Notepad can record the actual temperature of the bath from either the in-built indicator of an Isotech SITE model or from a variety of external instruments - see list. If the external instrument has two channels (e.g. Isotech TTI) then the unit under test may be connected to channel B for logging with CNP. Alternatively the value can be typed in from the keyboard. Similarly the calibration bath controller value, actual temperature, SITE indicator value or unit under test value may also be entered manually.

Then the manually entered data is combined with that gathered automatically and the resultant file can be opened in an external application such as Excel* for the preparation of reports, certificates etc.

Cal Notepad can :-

  • Record a Calibration Baths Controller Temperature
  • Record the Actual Baths temperature with either
  • A SITE model inbuilt indicator
  • A supported external indicator
  • Manual entry from the keyboard
  • Identify the Unit Under test with a serial number and have its output recorded
  • Export the log file to produce a complete report
  • Permit real time display of temperatures
  • Allow the calibration bath temperature to be changed from the PC
    • Temperature Programming
    • Program the bath to change temperature with time
    • Ramp to Temperature
    • Program the bath to ramp between limits ideal for liquid in glass thermometry
    • Open new chart for individual channels
    • Up to four charts in addition to the main screen

© 2008 Isothermal Technology Ltd. All rights reserved.