Because the 670SQ goes beyond the temperature range of oxide growth to the level at which the oxide
dissociates, the 670s is filled with a unique argon/oxygen mixture. A 2-metre length of low thermal EMF, high
temperature, screened cable is connected in the handle, via a strain-relieving transition, to the all pureplatinum
construction of the thermometer.
Gold-plated U-shaped terminals complete the cable construction, and the 670SQ is delivered in an elegant
soft lined carry-case of our own design.
A 670SQ is supplied only after a stabilising process which is complete when the reproducibility of RTPW is
within 0.0005°C after excursions to the extremes of its temperature range. Values of RTPW and WGA are
routinely provided with the 670SQ.
The 670SQ can be supplied with RTPW and WGA only or with full UKAS calibration. "With calibration" means
that you will get an Internationally accepted Fixed Point calibration. See databook 5 for details.
For best accuracy, recommended maximum measuring currents for the 670SQ are 1mA for the 25.5W (Ro)
and 0.5mA for the 100W Ro.
A comprehensive handbook and tutorial will help you get the very best performance and stability from your
670SQ.
The 670SQ 650mm long is our recommended SPRT offering ultra stability, vibration, shock, immersion and
self heating characteristics.
From the success of the original Model 670 SPRTs we have introduced new models into the 670 range offering
metal sheathed and low temperature models.