Gallium Fixed Point
Sostmann-Isotech Gallium Cell and Apparatus
Gallium Apparatus Simple, Automatic, Proven.
  DB1From Databook 1
  • 20 years of history without breakage
  • Automatically realise the plateau
  • 250mm immersion
  • Data Sheet
    Model 17402B Gallium Apparatus is not an adaptation from general purpose equipment, but is designed specifically to realize and maintain the Gallium Cell on the melt plateau, for calibration of thermometers on the International Temperature Scale of 1990.

    The Cell will realize the melt plateau of gallium (29.7646°C). It is furnished with a certificate of verification or optionally full UKAS calibration can be provided at extra cost.

    The apparatus will permit the gallium to melt coaxially over 12 to 16 hours. It is uniquely designed to also freeze the cell from the bottom up, which eliminates the danger of damage due to the expansion of gallium during freezing.

    1. The Cell contains about 0.5 kg of electronic grade gallium, 99.999999% pure. The gallium is contained within a resilient inner housing, which safely allows its solidus expansion, surrounded by an aluminium sleeve for longitudinal uniformity of temperature.

    2. Since the melting temperature of pure gallium is a recognized constant of nature (and a defining Fixed Point of ITS-90) recertification is normally never required.

    3. Model 17401 Gallium Cell may be used, without the thermal environment provided by Model 17402B Gallium Apparatus, in a well controlled bath. However, the advantages of automatic operation, convenience and cell protection, recommend the use of the Model 17402B environment in most cases.

    4. No external connections other than power are required.

    5. A completely automatic electronic control system provides a precise means for realization and maintenance of the plateau. The Apparatus can be turned on by a timer an hour before the laboratory day begins, the plateau utilized throughout the working day and the system recycled overnight. A thermal sink is provided which forces the gallium to refreeze upwards from the bottom (gallium expands when it freezes, requiring a specific freeze orientation to avoid rupturing the cell).

    6. Confidence is a major requirement in a standard. The Isothermal Gallium Cell and Apparatus have a long (20 year) history and have been successfully used in most National and Primary Laboratories world-wide. This is one of the main reasons for choosing the Isothermal Cell and Apparatus.

    7. The Cell and Apparatus are manufactured to the strict requirements of BS 5750 and ISO 9000. Each finished unit becomes part of the Isotech UKAS working Calibration Laboratory for two weeks or more, during which use the control circuits are set and checked for optimum performance. It is released only when Isotech is entirely satisfied with its performance.