| Company | Products | Lab | Technical Library | Training | What's New | Contact Us |

 

Furnaces for the Primary Laboratory
Medium Temperature Furnace Model 17703
Low Temp Furnace For Indium, Tin, Zinc and Aluminium
  DB1From Databook 1
  • Three Zone Control
  • Good Uniformity
  • Ten to Twenty Hour Plateau
  • Data Sheet
    The Isotech Medium Temperature Furnace is designed specifically to realize and maintain the freeze plateaux of Isotech Indium, Tin, Zinc and Aluminium Cells, for the calibration of thermometers on ITS-90. It can also be used with an insert, as an annealing facility of the highest order for SPRT’s or as a comparison calibration facility.

    A. The Medium Temperature Furnace will operate between 50°C and 700°C.

    B. The Medium Temperature Furnace is operated by 3 controllers in a master/slaves configuration which enables small temperature differences to be achieved along the Furnace. This is important when freezing cells, since the assumption made is that Cells freeze in concentric shells. This is true only if there is a small temperature gradient along the furnace. The controller resolution is 0.1°C.

    C. The recommended procedure for establishing a freeze plateau requires operator attention until the plateau is realized. Following that, the Furnace will maintain the plateau essentially automatically for a period of 10 to 20 hours, (longer if the heat flux from the Cell minimised).