| Company | Products | Lab | Technical Library | Training | What's New | Contact Us |

 

Secondary Laboratory
Temperature Controlled AC/DC Resistor
Std Resisitor

Wide range of values, temperature controlled

  DB2From Databook 2
  • High Stability
  • Low Temperature Coefficient
  • Wide Range of Values
Data Sheet
By temperature controlling an otherwise very stable resistor a performance close to the very best available World-wide can be achieved at a surprisingly low price. The resistor itself is oil filled and hermetically sealed.

The function of hermetic sealing is to eliminate the ingress of moisture and oxygen both of which play a role in the long term degradation of unsealed resistors. A further enhancement in both short and long term stability is achieved by oil filling. The oil also acts as a thermal conductor allowing the device to accept short periods of overload without degradation.

With accuracies of ±0.005% and a resistance range from 5 ohms to 3.3 megaohms and long term drift of less than 5ppm, these devices are virtually secondary standards that can be kept in a laboratory as references to calibrate other devices.

The Resistor is held in a temperature controlled environment heated to 30±0.1ºC other temperatures are available to special order. The heater requires 2 watts at 5V which can be supplied by a battery or an unregulated DC supply. In an ambient of 20ºC the Resistor’s heater will warm up in typically 30 minutes, and a LED shows when the temperature has been reached. A test pocket is provided so that the resistors’ temperature can be monitored if required.

0.1 ppm/month or better stability can be expected.

We can supply the value you choose ±0.1% between a minimum of 5 ohms and a maximum of 3.3 megaohms. However we bulk buy and keep in stock the following standard values: 10Ohm, 25Ohm, 100Ohm, 1000Ohm, 10,000Ohm.

For the highest quality traceability we recommend that the 456 be UKAS Certified.

© 2008 Isothermal Technology Ltd. All rights reserved.

Home Page Primary Lab Secondary Lab Industrial Calibration InfraRed Calibration